Technical Specifications for Transcutaneous Electrical Nerve Stimulator (TENS) Mode:
Waveform
Mono-phase square pulse wave
Pulse amplitude
Adjustable, 0~105mA peak at 1000 ohm Load each channel, 1mA/Step.
Pulse Width
Adjustable, from 50 to 300us microseconds, 10μS/step
Pulse Rate
Adjustable, from 1 to 150 Hz, 1 Hz/step
Burst (P1)
Burst rate: Adjustable, 0.5 ~ 5Hz; 0.1Hz/step Pulse width adjustable, 50~300μS Frequency fixed = 100 Hz
Normal (P2)
The pulse rate and pulse width are adjustable. It generates continuous stimulation based on the setting value.
Pulse Width Modulation (P3)
The pulse width is automatically varied in a cycle time. The pulse width is decreased from its original setting to 60% in setting cycle time, and then increased from 60% to its original setting in nest setting cycle time. In this program, pulse rate (1 to 150Hz), pulse width (50 to 300us) and cycle time (5 to 30 sec) are fully adjustable.
Pulse Rate Modulation (P4)
The pulse rate is automatically varied in a cycle time. The pulse rate is decreased from its original setting to 60% in setting cycle time, and then increased from 60% to its original setting in nest setting cycle time. In this program, pulse rate (1 to 150Hz), pulse width (50 to 300us) and cycle time (5 to 30 sec) are fully adjustable.
Technical Specifications for Interferential (IF) Mode:
Waveform
Bi-phase square pulse
Pulse amplitude
Adjustable, 0~70mA peak to peak at 1000 ohm Load each channel, 1mA/Step.
Pulse Rate
Channel 1 – Fundamental frequency: 4000 Hz fixed Channel 2 – Selectable frequency: 4001 to 4150 Hz
Interference frequency: 1 to 150 Hz.
Phase Width
125μs
P1
The pulse rate of the CH1 is fixed in 4000Hz; CH2 pulse rate is increased from 4001Hz to 4010Hz in a cycle time, and then decreased from 4010Hz to 4001Hz in nest setting cycle time. In this program, CH2 interference frequency is varied from 1Hz to 10Hz, cycle time (5 to 30 sec) is fully adjustable. CH 2 pulse rate=4000Hz+ Interference frequency
P2
The pulse rate of the CH1 is fixed in 4000Hz; CH2 pulse rate is increased from 4001Hz to 4150Hz in a cycle time, and then decreased from 4150Hz to 4001Hz in nest setting cycle time. In this program, CH2 interference frequency is varied from 1Hz to 150Hz, cycle time (5 to 30 sec) is fully adjustable. CH 2 pulse rate=4000Hz+ Interference frequency
P3
The pulse rate of the CH1 is fixed in 4000Hz; CH2 pulse rate is increased from 4080Hz to 4150Hz in a cycle time, and then decreased from 4150Hz to 4080Hz in nest setting cycle time. In this program, CH2 interference frequency is varied from 80Hz to 150Hz, cycle time (5 to 30 sec) is fully adjustable. CH 2 pulse rate=4000Hz+ Interference frequency
14P4
The pulse rate of the CH1 is fixed in 4000Hz; CH2 pulse rate is automatically varied in a cycle time. Interference frequency is increased from its original setting to 60% in setting cycle time, and then decreased from 60% to its original setting in nest setting cycle time. In this program, CH2 interference frequency (2 to 150Hz) and cycle time (5 to 30 sec) are fully adjustable.
CH 2 pulse rate=4000Hz+ Interference frequency